Short answer
Essentially, yes. A TDS or "ppm" meter measures electrical conductivity and multiplies it by an assumed factor: 0.5 on the 500 scale, 0.7 on the 700 scale, sometimes 0.64 [1]. The same 1.0 mS/cm solution reads 500 ppm on one meter and 700 on another. Work in EC (mS/cm), or always state the scale. Established.
Why
Dissolved salts carry current; more ions, more conductivity. That is what the probe measures, corrected to 25 °C [1].
Converting conductivity to dissolved solids needs a factor, and no single factor is right. Different salts conduct differently per gram, and for natural waters the true factor rises from about 0.65 at 1,000 µS/cm to about 0.82 at 10,000 µS/cm [1]. Meter makers just pick one.
EC also says nothing about which ions are present [2]. Sodium chloride has EC too, so a reading tells you strength, not balance.
The numbers
| EC meter | TDS / ppm meter | |
|---|---|---|
| What the probe measures | Conductivity | Conductivity |
| Displays | mS/cm or µS/cm | EC × 0.5, 0.64 or 0.7 |
| 1.0 mS/cm reads | 1.0 | 500, 640 or 700 ppm |
| 1.413 mS/cm standard reads | 1.413 | ~707 (500) or ~990 (700) |
| Comparable across meters | Yes | Only on the same scale |
| Tells you which nutrients | No | No |
Verdict: same sensor, different label; choose a meter that can show EC and use that.
What to do
- Check your meter's scale in the manual or settings.
- Switch the display to EC if it allows.
- Calibrate against a 1.413 mS/cm standard.
- Convert any ppm feed chart to EC using its stated scale.
- Measure your water first; feed charts usually mean total EC.
Common mistakes
- Following a 700-scale chart with a 500-scale meter and overfeeding by 40%.
- Comparing ppm numbers between growers on different meters.
- Reading TDS ppm as ppm of nitrogen.
- Using a meter with no temperature compensation.