Short answer
Do a salt test: seal the hygrometer for 12–24 hours with a damp slurry of table salt, which holds the air at about 75 % RH at room temperature [1]. The difference from 75 is your offset. For a second point, magnesium chloride gives about 33 % [1]. Established physics.
Why
A saturated salt solution in a sealed space settles at a fixed relative humidity set by the salt and the temperature. These "humidity fixed points" are a standard reference for calibrating humidity sensors [1]. Sodium chloride sits near 75 % across normal room temperatures, which makes table salt a cheap, reliable check.
Cheap RH sensors drift 3–5 % RH. Every VPD calculation uses RH [2], so a sensor reading 5 % low can have you running a late-flower room far more humid than you think.
That matters because humidity is an established driver of cannabis bud rot and powdery mildew [3]. A wrong number is worse than no number if you trust it.
The numbers
| Reference | RH at about 25 °C (77 °F) |
|---|---|
| Saturated sodium chloride (table salt) | ~75 % [1] |
| Saturated magnesium chloride | ~33 % [1] |
| Typical cheap sensor drift | 3–5 % RH |
| Acceptable offset to just correct for | within about ±3 % |
| Replace if | off by more than ~5 % or changing between tests |
What to do
- Put a quarter cup of plain salt in a cap or dish; add water until it's a damp slurry, with undissolved salt visible.
- Seal it and the hygrometer in a zip bag or jar. Keep the slurry off the sensor.
- Leave it 12–24 hours at a steady room temperature, out of sun.
- Read it. Note the offset (for example, reads 71 % = 4 % low).
- Adjust the meter if it allows; otherwise add the offset mentally and label it.
- Optional: repeat with magnesium chloride for a low point.
- Retest every few months, and after any humid spell.
Common mistakes
- A wet salt solution with no undissolved crystals.
- Reading after an hour instead of 12–24.
- Testing in a room with big temperature swings.
- Trusting a sensor near the light or in still air.
- Correcting a drifting sensor forever instead of replacing it.